Tomtec Elutrix and Harvex rapid SPE and Elution
Tomtec Elutrix SPE system w/ Harvex unit. more info
SPExpress Sample Loading
Ninety-six discreet flow paths are formed by the Harvex to insure there is no cross contamination. Vacuum is applied to the top of the card to draw reagents through the card. Simply change the reservoir or microtiter plate on the sample loading tray to activate, load sample, and wash the cards. Samples can be loaded from microtiter plates or test tubes in the 96 well format. No dry-down or reconstitution steps are required. Complete preparation of a 96 well card can be performed in less than 5 minutes.
Sample Elution
EmporeĀ® cards are eluted directly to the detector (Mass Spec, UV...) using the Elutrix. Each zone is individually sealed, and a switching valve is activated to pass the mobile phase through the elution zone. The 96 zones on the card can be processed using four pre-programmed patterns, or in any order chosen by the customer utilizing up to 25 custom patterns. The Elutrix and the detector are synchronized via hardware signals to ensure accurate timing when processing the sample. The system is designed for running single cards or unattended operation with as many as thirty cards using the stacking cassettes.
McKinley Scientific is a LabX Sponsor.
Please visit their website at: www.mckscientific.com
|